M550129-03 Rev G www.programmablepower.com SG Series IEEE 488.2/RS232 and Ethernet Options Programming Manual
Contents SG Series Programming xii M550129-03 Rev G This page intentionally left blank.
SCPI Command Operation SG Series Programming 6-38 M550129-03 Rev G Command Description SCPI :SELected Causes the presently selected sequen
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-39 Command Description SCPI :NAME “sequence name” Performs one of two possible
SCPI Command Operation SG Series Programming 6-40 M550129-03 Rev G Command Description SCPI :STATe? Returns the present state for the selecte
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-41 6.13.4 SGI MEASURE SCPI Command Subsystem The POWer query was added to the Measu
SCPI Command Operation SG Series Programming 6-42 M550129-03 Rev G Command Description SCPI :ARM <boolean> Readies the last loaded seque
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-43 // Turn on the unit. *CLS // clear the unit to its power-on default settin
SCPI Command Operation SG Series Programming 6-44 M550129-03 Rev G 6.14.4 Hardware Trigger Example This example applies only to units with the Ex
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-45 SOUR:CURR 5.0 // program output current to 5.0 A. SOUR:CURR:RAMP 25.0 30.0
SCPI Command Operation SG Series Programming 6-46 M550129-03 Rev G SOUR:CURR? // confirm power-on initial current setting. SOUR:VOLT? // confirm
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-47 PROG:DEF 1, VIMODE,3,4,11,10 //go to 3 volts, 4 amps, 11 volts ovp, for 10
M550129-03 Rev G 1-1 SECTION 1 OVERVIEW 1.1 INTRODUCTION This manual provides instructions for full remote programming control and monitoring from
SCPI Command Operation SG Series Programming 6-48 M550129-03 Rev G EXAMPLE 2 Issue *RST to establish a known state. PROG:NAME “SEQ2” PROG:MALL DEF
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-49 EXAMPLE 3 Issue *RST to establish a known state. In this third example, redefine
SCPI Command Operation SG Series Programming 6-50 M550129-03 Rev G OUTP:STAT ON PROG:STAT RUN The output of the supply for Example 3 should look
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-51 PROG:DEF 9, STOP //stop running the sequence w
SCPI Command Operation SG Series Programming 6-52 M550129-03 Rev G The formula for charging is )1(10RCte−− . Issue *RST to establish a known stat
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-53 volts in 0.1 sec) PROG:DEF 17, RAMPTOV, 7.77, 7.98, 5, 20, 0.1 (ramp from 7
SCPI Command Operation SG Series Programming 6-54 M550129-03 Rev G PROG:DEF 8, RAMPTOV, 3.82, 3.46, 5, 20, 0.1 //ramp from 3.82 volts down to 3.4
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-55 PROG:NAME “Charge” OUTP:STAT ON PROG:STAT RUN The output for Example 5 should lo
SCPI Command Operation SG Series Programming 6-56 M550129-03 Rev G EXAMPLE 6 This example shows how to make an infinite loop for the creation of a
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-57 To stop the execution of the square wave sequence, issue the STOP state command:
Overview SG Series Programming 1-2 M550129-03 Rev G This page intentionally left blank.
SCPI Command Operation SG Series Programming 6-58 M550129-03 Rev G EXAMPLE 7 This example shows how to make a definite length square wave (as oppo
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-59 OUTP:STAT ON PROG:STAT RUN The power supply output for Example 7 should look som
SCPI Command Operation SG Series Programming 6-60 M550129-03 Rev G This page intentionally left blank.
M550129-03 Rev G 7-1 SECTION 7 CALIBRATION CAUTION Please refer to the power supply operation manual for further information before performing cal
Calibration SG Series Programming 7-2 M550129-03 Rev G be issued once calibration procedures have been performed. 7.2 SETUP FOR CALIBRATION This se
SG Series Programming Calibration M550129-03 Rev G 7-3 7.3 VOLTAGE PROGRAMMING CALIBRATION (ETHERNET) STEP DESCRIPTION 1. Issue a *CLS command.
Calibration SG Series Programming 7-4 M550129-03 Rev G 13. Enter the actual output voltage value of the third calibration point: CAL:OUTP:VOLT:FIVE
SG Series Programming Calibration M550129-03 Rev G 7-5 7.4 VOLTAGE PROGRAM GAIN/OFFSET AND MEASUREMENT READBACK CALIBRATION (ETHERNET, GPIB) NOTE:
Calibration SG Series Programming 7-6 M550129-03 Rev G 11. Program the DIA to unlock the non-volatile memory for calibration value storage: CAL:UNL
SG Series Programming Calibration M550129-03 Rev G 7-7 7.6 CURRENT PROGRAMMING CALIBRATION (ETHERNET) Preparation: 1. Power down the Unit Under
M550129-03 Rev G 2-1 SECTION 2 IEEE 488.2 GPIB/RS232 FEATURES, FUNCTIONS AND SPECIFICATIONS 2.1 INTRODUCTION This section introduces the feature
Calibration SG Series Programming 7-8 M550129-03 Rev G 11. Program the output of the third calibration point by sending the following command strin
SG Series Programming Calibration M550129-03 Rev G 7-9 7.7 CURRENT PROGRAMING GAIN/OFFSET AND MEASUREMENT READBACK CALIBRATION (ETHERNET, GPIB) NO
Calibration SG Series Programming 7-10 M550129-03 Rev G 10. Program the Digital Interface Adapter (DIA) to calculate the current readback calibrati
SG Series Programming Calibration M550129-03 Rev G 7-11 7.8 ANALOG PROGRAM ADJUSTMENT This sections applies only to units with the Isolated Inte
Calibration SG Series Programming 7-12 M550129-03 Rev G Figure 7-1. Potentiometer Locations STANDARD CALIBRATION ADJUSTMENT R35 = VPRES (1mA) Adjus
SG Series Programming Calibration M550129-03 Rev G 7-13 Preparation: 1. Remove prime AC power. 2. Connect the power supply to a precision curren
Calibration SG Series Programming 7-14 M550129-03 Rev G 7.8.1 Adjustment for Current Mode STEP DESCRIPTION 1. Set the SGI Series unit to operate
SG Series Programming Calibration M550129-03 Rev G 7-15 7. Verify the unit is set to 100% ± 0.8% of full-scale output current. If necessary, adjus
Calibration SG Series Programming 7-16 M550129-03 Rev G 6. Verify that the unit is set to 100% ± 0.25% of full-scale output voltage. If necessary a
SG Series Programming Calibration M550129-03 Rev G 7-17 This page intentionally left blank.
IEEE/GPIB Features, Functions, Specifications SG Series Programming 2-2 M550129-03 Rev G 2.4 READBACK FUNCTIONS • Actual measured voltage and cur
M550129-03 Rev G 8-1 SECTION 8 SCPI STATUS IMPLEMENTATION
SCPI Status Implementation SG Series Programming 8-2 M550129-03 Rev G This page intentionally left blank.
M550129-03 Rev G 8-3 Various fault events must be enabled by the protection enable register before they are recorded in the fault register. Bit He
This page intentionally left blank.
M550129-03 Rev G Index-1 INDEXA Accuracy programming, Ethernet, 4-4 readback, Ethernet, 4-5 Analog Program Adjustment, 7-11 atures and Functions Et
Index SG Series Programming Index-2 M550129-03 Rev G ISOLATION output signal, 5-11 POLARITY output signal, 5-11 SENSE output signal, 5-11 SYNC outpu
SG Series Programming Index M550129-03 Rev G Index-3 Ramping Description, 6-16, 6-25 Readback accuracy, Ethernet, 4-5 resolution, Ethernet, 4-5 Read
About AMETEK AMETEK Programmable Power, Inc., a Division of AMETEK, Inc., is a global leader in the design and manufacture of precision, programmable
M550129-03 Rev G 3-1 SECTION 3 IEEE 488.2 GPIB/ RS232 CONFIGURATIONS AND REMOTE PROGRAMMING 3.1 REAR PANEL This section provides illustrations of
This page intentionally left blank. ii
Important Safety Instructions Before applying power to the system, verify that your product is configured properly for your particular application.
This page intentionally left blank. iv
Product Family: SG Series IEEE 488.2/RS232 and Ethernet Options Warranty Period: Five Years WARRANTY TERMS AMETEK Programmable Power, Inc. (“AMETEK”)
vi This page intentionally left blank.
IEEE/RS232 Configuration and Remote Programming SG Series Programming 3-2 M550129-03 Rev G Figure 3-2. SG Unit with GPIB/RS232 Option 1 – RS232
SG Series Programming IEEE/RS232 Configuration and Remote Programming M550129-03 Rev G 3-3 3.2 RS232 SETUP PROCEDURE This procedure is a quick ref
IEEE/RS232 Configuration and Remote Programming SG Series Programming 3-4 M550129-03 Rev G 6. Establish communication. 7. Test the communication i
SG Series Programming IEEE/RS232 Configuration and Remote Programming M550129-03 Rev G 3-5 3.3.1 Configuration Switch The DIP switch (may be 8-pin
IEEE/RS232 Configuration and Remote Programming SG Series Programming 3-6 M550129-03 Rev G 3.3.2 Remote/Local Selection Set the rear panel Remote/
SG Series Programming IEEE/RS232 Configuration and Remote Programming M550129-03 Rev G 3-7 Table 3-2. Remote Power-on Conditions Condition Default
IEEE/RS232 Configuration and Remote Programming SG Series Programming 3-8 M550129-03 Rev G 3.3.3 Power-On GPIB Service Request (PON SRQ) Selection
SG Series Programming IEEE/RS232 Configuration and Remote Programming M550129-03 Rev G 3-9 3.4 REMOTE PROGRAMMING VIA RS232 The RS232 interface ope
IEEE/RS232 Configuration and Remote Programming SG Series Programming 3-10 M550129-03 Rev G This page intentionally left blank.
M550129-03 Rev G 4-1 SECTION 4 ETHERNET INTERFACE FEATURES, FUNCTIONS AND SPECIFICATIONS 4.1 INTRODUCTION This section covers the Remote Programmin
Ethernet Features, Functions and Specifications SG Series Programming 4-2 M550129-03 Rev G INTERNET OR LAN CONNECTION To operate your Sorensen Ether
SG Series Programming Ethernet Features, Functions and Specifications M550129-03 Rev G 4-3 4.2.2 Programmable Functions • Output voltage and curre
Ethernet Features, Functions and Specifications SG Series Programming 4-4 M550129-03 Rev G 4.3.2 Ethernet Configuration Factory Defaults PARAMETER
SG Series Programming Ethernet Features, Functions and Specifications M550129-03 Rev G 4-5 4.3.5 Readback Resolution SGA SGI Voltage 0.002% of f
Ethernet Features, Functions and Specifications SG Series Programming 4-6 M550129-03 Rev G This page intentionally left blank.
M550129-03 Rev G 5-1 SECTION 5 ETHERNET CONFIGURATION AND REMOTE PROGRAMMING 5.1 REAR PANEL This section provides illustrations of the SG power su
Ethernet Configuration and Remote Programming SG Series Programming 5-2 M550129-03 Rev G Figure 5-2. SG Rear Panel with Ethernet/RS232 Options 1 –
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-3 Figure 5-3. SGI Rear Panel with Ethernet/RS232 Options (4
Ethernet Configuration and Remote Programming SG Series Programming 5-4 M550129-03 Rev G NOTE: The SG Ethernet Option has been designed and tested t
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-5 Interface differs in that its banner shows “SGA” and it doe
Ethernet Configuration and Remote Programming SG Series Programming 5-6 M550129-03 Rev G b. Click open Network Connections. (For XP, if in the Categ
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-7 d. In the ASCII Setup window in the ASCII Sending section,
Ethernet Configuration and Remote Programming SG Series Programming 5-8 M550129-03 Rev G 5.2.5 Configuration Switch Use the DIP switch (will have e
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-9 5.2.6 Remote/Local Selection Set the rear panel Remote/Loc
Ethernet Configuration and Remote Programming SG Series Programming 5-10 M550129-03 Rev G Table 5-2. Remote Mode Power-on Conditions Condition Defa
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-11 Table 5-3. External User Control Signal Connector Pinout
Ethernet Configuration and Remote Programming SG Series Programming 5-12 M550129-03 Rev G Figure 5-8. Example of Open Collector, TTL Input, and R
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-13 5.4 PROGRAMMING/COMMUNICATION VIA ETHERNET With the Ether
Ethernet Configuration and Remote Programming SG Series Programming 5-14 M550129-03 Rev G 5.5 ETHERNET WEB PAGES, OVERVIEW The layout of each of th
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-15 Figure 5-11. Login Window Once you have logged in, the L
Ethernet Configuration and Remote Programming SG Series Programming 5-16 M550129-03 Rev G 5.5.1 HOME This is the default, information-only page. It
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-17 • IP Address: your power supply’s address actually in use
Ethernet Configuration and Remote Programming SG Series Programming 5-18 M550129-03 Rev G To change: Type the new name (15 characters maximum) in t
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-19 update (or make all desired changes before clicking Apply)
M550129-03 Rev G vii CONTENTS SECTION 1 OVERVIEW ...1-1 1.1 Introduction...
Ethernet Configuration and Remote Programming SG Series Programming 5-20 M550129-03 Rev G Primary: Obtain an IP Address Automatically (DHCP) and Aut
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-21 5.5.3 SETTINGS The Settings page is available to users wh
Ethernet Configuration and Remote Programming SG Series Programming 5-22 M550129-03 Rev G FAULT indicator: highlighted red if fault has occurred OUT
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-23 Figure 5-15. Alert Message for Save Settings SCPI COMMAN
Ethernet Configuration and Remote Programming SG Series Programming 5-24 M550129-03 Rev G 5.5.4 STATUS The Status page displays updated information
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-25 Last Calibration Date: displays the date that the power su
Ethernet Configuration and Remote Programming SG Series Programming 5-26 M550129-03 Rev G 5.5.5 POWER The Power page (only in SGI) displays updated
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-27 Stop Current: the programmed current setting when you exit
Ethernet Configuration and Remote Programming SG Series Programming 5-28 M550129-03 Rev G 5.5.6 PRESETS The Presets page (only in SGI) displays sto
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-29 • Over Volt: click in the Over Volt field and input a new
Contents SG Series Programming viii M550129-03 Rev G SECTION 4 ETHERNET INTERFACE FEATURES, FUNCTIONS AND SPECIFICATIONS...
Ethernet Configuration and Remote Programming SG Series Programming 5-30 M550129-03 Rev G 5.5.7 SECURITY Accessible only if you have Administrative
SG Series Programming Ethernet Configuration and Remote Programming M550129-03 Rev G 5-31 ADD NEW USER Accessible from the Security Page by clicking
Ethernet Configuration and Remote Programming SG Series Programming 5-32 M550129-03 Rev G EDIT EXISTING USER Accessible from the Security Page by cl
M550129-03 Rev G 6-1 SECTION 6 IEEE 488.2 GPIB/RS232/ETHERNET AND SCPI COMMAND OPERATION 6.1 INTRODUCTION This section describes the operation of t
SCPI Command Operation SG Series Programming 6-2 M550129-03 Rev G Table 6-1. SCPI Status Byte Bit Hex Value Description 0 0x01 Not used. 1 0x02 P
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-3 6.2.2 Standard Event Status Register (ESR) Read the Standard Event Status Registe
SCPI Command Operation SG Series Programming 6-4 M550129-03 Rev G requests based on the Protection Event Status Register, program both the Protect
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-5 6.2.4 Operation Status and Questionable Status Registers The Operation Status and
SCPI Command Operation SG Series Programming 6-6 M550129-03 Rev G Error Code Description 201 Unexpected warm boot This error means that the DIA p
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-7 Error Code Description -330 Self-test failed A self-test failure has occurred. -3
SG Series Programming Contents M550129-03 Rev G ixSECTION 6 IEEE 488.2 GPIB/RS232/ETHERNET AND SCPI COMMAND OPERATION ...
SCPI Command Operation SG Series Programming 6-8 M550129-03 Rev G 6.2.6 Serial Poll Operation Performing a serial poll will not modify the Status
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-9 6.3.2 Units The SGA/SGI power supplies will accept the following units as suffixe
SCPI Command Operation SG Series Programming 6-10 M550129-03 Rev G 6.4 IEEE 488.2 COMMON COMMAND SUBSYSTEM The following commands are common to
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-11 Command Description *SRE <0+NR1> Sets the value of the Service Request Ena
SCPI Command Operation SG Series Programming 6-12 M550129-03 Rev G 6.5 SOURCE SCPI COMMAND SUBSYSTEM This section first presents a tree summary o
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-13 :PROTection? [:LEVel] <NRf> [:LEVel?] :TRIPped? :STATe? :CLEar
SCPI Command Operation SG Series Programming 6-14 M550129-03 Rev G Command Description SCPI :ALL? Returns the ramping status of all channe
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-15 Command Description SCPIin millivolts. :PROTection? Returns the set overvol
SCPI Command Operation SG Series Programming 6-16 M550129-03 Rev G Command Description SCPI :TRIGgered <NRf> <NRf> Sets the ou
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-17 6.6 MEASURE SCPI COMMAND SUBSYSTEM This section first presents a tree summary of
Contents SG Series Programming x M550129-03 Rev G 6.12.2 CALIBRATION SCPI Command Reference...6-29 6.13 SGI-Uni
SCPI Command Operation SG Series Programming 6-18 M550129-03 Rev G 6.7 OUTPUT SCPI COMMAND SUBSYSTEM This section first presents a tree summary o
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-19 Command Description SCPI :FOLD <0|1|2> Sets the foldback (program down
SCPI Command Operation SG Series Programming 6-20 M550129-03 Rev G 6.8.2 STATUS SCPI Command Reference The letter “C” in the “SCPI” column means
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-21 6.9 SYSTEM SCPI COMMAND SUBSYSTEM This section first presents a tree summary of
SCPI Command Operation SG Series Programming 6-22 M550129-03 Rev G 6.9.2 SYSTEM SCPI Command Reference The letter “C” in the “SCPI” column means
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-23 Command Description SCPI :DHCPMODE? Returns 1 if DHCP Mode is enabled in t
SCPI Command Operation SG Series Programming 6-24 M550129-03 Rev G Command Description SCPI :TERM <NRf> Sets the return string termina
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-25 6.10 HTRIGGER SCPI COMMAND SUBSYSTEM This section applies only to units with the
SCPI Command Operation SG Series Programming 6-26 M550129-03 Rev G 6.11 TRIGGER SCPI COMMAND SUBSYSTEM This section describes the programming sof
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-27 6.12 CALIBRATION SCPI COMMAND SUBSYSTEM Note: See Section 7 for calibration pr
SG Series Programming Contents M550129-03 Rev G xiTable 6-1. SCPI Status Byte...
SCPI Command Operation SG Series Programming 6-28 M550129-03 Rev G :OUTPut :CURRent :CALCulate :DAC <NR1> :FIVEPoint <1|2|3|4|
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-29 6.12.2 CALIBRATION SCPI Command Reference The letter “C” in the “SCPI” column me
SCPI Command Operation SG Series Programming 6-30 M550129-03 Rev G Command Description SCPI :CALCulate Calculates the value of the gain a
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-31 Command Description SCPI :VOLTage Calibrate Output Voltage subsystem N
SCPI Command Operation SG Series Programming 6-32 M550129-03 Rev G 6.13 SGI-UNIQUE COMMANDS This section deals with the SGI power supply’s unique
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-33 6.13.2 SGI SOURCE SCPI Command Subsystem The Power command and its corresponding
SCPI Command Operation SG Series Programming 6-34 M550129-03 Rev G 6.13.3 SGI PROGRAM SCPI Command Subsystem The PROGRAM subsystem, only availabl
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-35 :SAVe :SELected :ALL :STATe <value> :STATe? PROGRAM SCPI COMMAND RE
SCPI Command Operation SG Series Programming 6-36 M550129-03 Rev G Command Description SCPI :DEFine <step#>, RAMPTOC, <volts>, &l
SG Series Programming SCPI Command Operation M550129-03 Rev G 6-37 Command Description SCPI :DEFine <step#>, LOOP,<count> Programs